EasyManuals Logo

Pentair AUTOTROL LOGIX 742 User Manual

Pentair AUTOTROL LOGIX 742
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
Installer manual Autotrol Magnum 293 - 298 Logix 742-762 - Troubleshooting
Ref. MKT-IM-012 / D - 03.02.2022 91 / 110
Issue Cause Solution
Loss of water
pressure.
Fouled resin bed due to iron
accumulation.
Clean Magnum control valve and
mineral bed with cleaner.
Slots in riser pipe or laterals are filled
with resin fines.
Inspect and clean distributor pipe
slots as needed.
Salt in water to
service after
regeneration.
The injector is too small for the
system size.
Install the correct injector.
Brine draw time excessively long due
to low water pressure.
Increase water pressure above
172kPa minimum.
Restricted drain line. Remove drain line restriction.
Insufficient rinse volume. Increase slow rinse time, fast
rinse time or both.
Plugged injector and/or injector
screen.
Inspect and clean the injector
and/ or injector screen assembly.
10.3 Magnum valve cartridge
The following procedure provides sequential troubleshooting steps to isolate a suspect cartridge.
Info
The dynamic pressure applied to the valve must be greater than 172.4 kPa at all times
when performing the following tests.
If a cartridge needs to be removed, follow the removal procedure in section Cartridge
replacement [
Page85].
#2 Rinse cartridge
#3 Outlet cartridge
#4 Inlet cartridge
#1 Drain cartridge
There are four symptoms that may require a cartridge to be removed and inspected or replaced.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pentair AUTOTROL LOGIX 742 and is the answer not in the manual?

Pentair AUTOTROL LOGIX 742 Specifications

General IconGeneral
BrandPentair
ModelAUTOTROL LOGIX 742
CategoryControl Unit
LanguageEnglish

Related product manuals