LIBRARIES
PerkinElmer Life and Analytical Sciences 117
Name
This is the name of the Alpha Beta Standards set. To add a new standards set to
the library, click the Add button. A blank line appears in the table. Type in the
desired name for the Alpha Beta Standard Set.
Beta:LL
This is the lower limit of the counting region for the Beta nuclide, in keV.
Note: This field is for information only in this library.
Beta:UL
This is the upper limit of the counting region for the Beta nuclide, in keV.
Note: This field is for information only in this library.
Alpha:LL
This is the lower limit of the counting region for the Alpha nuclide, in keV.
Note: This field is for information only in this library.
Alpha:UL
This is the upper limit of the counting region for the Alpha nuclide, in keV.
Note: This field is for information only in this library.
Discriminator Type
Select either Automatic or Manual from the drop-down list. Selecting Automatic
allows the instrument to determine the pulse decay discriminator range. Selecting
Manual allows you to enter the upper and lower limits for the pulse decay
discriminator.
Note:When using Automatic mode, each pure standard source
must generate a minimum of 50,000 CPM (Counts Per Minute).
There is no activity requirement when using Manual mode.
Discriminator Range: LL
If you selected Manual as the Discriminator Type, you may enter a value between
0 and 255 for the lower limit of the Pulse Decay Discriminator Range.
Discriminator Range: UL
If you selected Manual as the Discriminator Type, you may enter a value between
0 and 255 for the upper limit of the Pulse Decay Discriminator Range.