CHAPTER 6
118 PerkinElmer Life and Analytical Sciences
Number of Points
Enter the number of points you would like to use to manually generate the Alpha
and Beta misclassification curves. This is not required if Automatic is chosen in the
Discriminator Type. Ten points are used if Automatic is chosen. Up to 15 points
can be used if Manual is chosen.
Discriminator Computed
This field represents the optimal Pulse Decay Discriminator value, as determined
by the instrument.
Discriminator in Use
This value represents the Pulse Decay Discriminator value currently in use.
Count Mode
Normal is the Count Mode used to count the Alpha and Beta Standards.
Note: This field is for information only.
Coincidence Time
This field represents the Coincidence Time of 18 ηs used when counting the Alpha
Beta Standards Assay.
Note: This field is for information only in this library.
Delay Before Burst
This field represents the Delay Before Burst time of 75 ηs used when counting the
Alpha Beta Standards Assay.
Note: This field is for information only in this library.
Date Counted
This field represents the date on which the Alpha Beta Standard Assay was
counted.
Note: This field is for information only.
Time Count Ended
This field represents the time at which the Alpha Beta Standard Assay was
completed.
Note: This field is for information only.
Add
Click this button to add a line to the table and enter a new Alpha Beta Standard
Set.