ADVANCED FEATURES
PerkinElmer Life and Analytical Sciences 139
To count samples using the SPC Decay mode:
1. Load your sample(s) into a cassette. Attach the Priostat flag in the reset
position to the cassette.
2. Select SPC Decay from the Run menu. The Sample Priostat window is
displayed.
3. Select Count from the Run menu. Sample counting commences immediately
after the instrument identifies the Priostat flag.
4. To stop the counting of a sample, select Stop from the Run menu.
5. Select Next Sample followed by Count from the Run menu if you would like to
count additional samples.
6. Stop the sample counting procedure by selecting End Sample Priostat from the
Run menu.