CHAPTER 8
144 PerkinElmer Life and Analytical Sciences
Reverse Region
To optimize the counting region settings such that they take into account your
sample’s chemical environment, Reverse Region compensation should be used.
Reverse Region determines region settings using a factor based on the tSIE and
the observed spectral endpoint of the sample.
Figure 8-5 Sample Priostat Window