Battery Related Problems
Check Settings INOP occurs during normal operation,
indicating a possible monitor software
problem
Check the monitor and patient settings
before you resume monitoring. If the
settings are unexpected, there may be a
problem with the monitor software.
1 Silence the INOP.
2 Load the User Defaults.
If this is unsuccessful, try loading the
Factory Default and reconfigure the
monitor in Configuration Mode, and save
the new settings in the User Defaults.
If the INOP persists, there is an unresolved
software problem. Report the problem to
factory support.
INOP occurs after a software upgrade,
indicating a possible incomplete or
unsuccessful upgrade
Clone the correct settings via the Support
Tool.
Check TI Config INOP occurs after a software upgrade or
option change, indicating that the monitor
settings need to be updated to support the
software feature "NST Report"
Check the setting item Setup NST Report,
Guideline. If set to "none", select the
appropriate NST Report Guideline. If no
Guideline is available, clone updated
settings via the Support Tool (e.g. settings
based on an initial configuration file for the
installed software revision, plus customer
specific adaptations, if required).
Internal.Comm.Malf Main CPU Board defective Replace the Main CPU Board.
Settings Malfunct Problem during cloning process Reclone the configuration file.
Memory space in which the settings are
stored has been corrupted
Reclone the configuration file. This will
reload the memory space.
Main CPU Board defective
Replace the Main CPU Board.
INOP Message Possible Causes of Failure Failure Isolation and Remedy
Symptoms Causes of Failure Failure Isolation and Remedy
Battery symbol is not displayed The monitor is not configured for
battery operation.
Make sure a system interface board with
battery capability is installed.
Batt Empty INOP tone, battery
LED flashes
During this INOP, alarms
cannot be paused or switched
off
The estimated remaining battery-
powered operating time of the battery
is 10 minutes.
Recharge the battery immediately.
If the condition persists, this INOP is re-
issued two minutes after you acknowledged
it.
Batt Incompat INOP tone The indicated battery cannot be used
with this monitor.
Replace with the correct battery (M4605A).