3 Installation Instructions
37
Configuring the Equipment Label
OB TraceVue/IntelliSpace Perinatal requires a unique equipment label. In OB TraceVue/IntelliSpace
Perinatal it is possible to prevent connection to monitors with specific equipment labels by the means
of a filtering mechanism. For more details, see the OB TraceVue/IntelliSpace Perinatal Instructions
for Use.
1 Select the Bed Label screen element to call up the Bed Info menu.
2 Select Equipment Label to call up the on-screen keyboard.
3 Enter the system identifier.
Configuring SmartKeys
Check that the configured SmartKeys are suitable. Configure the SmartKeys preferred by the
institution from a global list of Global Smart Keys. The global list of SmartKeys is stored as a unique
monitor setting in the monitor configuration. See the section Configuring Global SmartKeys in the
Configuration Guide for details on how to configure the global SmartKey list.
USB/PS/2 Keyboard/Mouse
1 Switch off the monitor before connecting any USB/PS/2 compatible device.
2 Connect the USB/PS/2 connector to the USB/PS/2 Interface board in the monitor at the slot
indicated by the appropriate symbol.
3 The default keyboard language setting for all initial configurations is "US".
4 To configure the keyboard language manually, go to Service Mode, select Main Setup, Hardware,
Keyboard and then select the proper language. Please note that this setting does not clone.