7 Testing and Maintenance
87
4
The expected test results depend on the relay contact used. Please check that the correct relay
activity is initiated during alarm condition.
Reporting of Test Results
Philips recommends all test results are documented in accordance with local laws. Authorized Philips
personnel report the test result back to Philips in a service record. While hospital personnel
(biomedical engineers or technicians) do not need to report results to Philips, Philips recommends that
they record and store the test results in accordance with local laws.
The following table lists what to record after completing the tests in this chapter. Record the results in
the empty column in the following table.
The following is a guide as to what your documentation should include:
• Identification of the testing body (for example, which company or department carried out the
tests).
• Name of the person(s) who performed the tests and the concluding evaluation.
• Identification of the device(s) and accessories being tested (serial number, etc.).
• The actual tests (incl. visual inspections, performance tests, safety, and system tests) and
measurements required.
• Date of testing and of the concluding evaluation.
• A record of the actual values of the test results, and whether these values passed or failed the tests.
• Date and confirmation of the person who performed the tests and evaluation.
The device under test should be marked according to the test result: passed or failed.
Test Expected test results
Multi-Port Nurse Call Connector Test Correct relay activity is initiated during alarm
condition (pass/fail)