6 Connection to a Network
54
Configuring the MIB/RS232 I/O Board Data Export Setting
The configuration of a specific MIB/RS232 port can be viewed in Configuration Mode and altered in
Service Mode.
Data Out can be configured up to two times (on two or more MIB/RS232 boards). Note
that only the first MIB/RS232 port configured to
Data Out (i.e. the first one to receive a request)
provides wave export. A second MIB/RS232 port configured to
Data Out will only export numerics.
1 To alter the configuration of an MIB/RS232 port select Main Setup, Hardware, Data Export 1, or
Data Export 2, and select the required setting.
2 Exit Configuration Mode. You do not need a password to return to Monitoring Mode.
NOTE
Be aware that if you change a port assignment, this assignment is not reset upon boot up. If the MIB/
RS232 board is removed and replaced with a different type of board, the settings are deleted. If the
MIB/RS232 board is then refitted, you must reconfigure the MIB/RS232 port. The configuration of
MIB/RS232 is not cloned between monitors.
AutoSpeed Transport protocol with baudrate negotiation, based on the
IrDA protocol.
Fix 19200 Transport protocol with a fixed baudrate of 19.200 baud.
Fix 115200 Transport protocol with a fixed baudrate of 115.200 baud.