11 Upgrades
215
Installing Upgrade Options
This section covers how to install the options.
Option B71
To install the noninvasive blood pressure upgrade:
1 Remove the top cover assembly as described in “Removing the Top Cover Assembly” on
page 141.
2 Remove the front end assembly (see “Removing the Front End Assembly” on page 200).
3 Remove the connector block frame from the front end assembly (see “Disassembling the Front
End Assembly” on page 202). Leave the fetal sensor socket connector block in place.
4 Take the new noninvasive blood pressure socket. Locate the metal pin and the plastic stud on the
underneath of the noninvasive blood pressure socket in the corresponding holes in the bus master
board.
J70 System Interface, 1 x RS232
port and 1 x LAN port.
RS232/LAN I/O card
assembly
M2703-67501 "Plug & Play" interface 1
SJ3 Rel. J.3 Software Upgrade Fetal Monitoring
Documentation DVD-
ROM
453564491101 -1
Instructions for Use 453564490751 Appropriate localized version 1
FM20/30 Upgrade Options
Option
Number
Option Adds
Parts included with Option
Description Part Number Contents/Comments Qty