10 Repair and Disassembly
174
Removing the Paper Sensor Assembly
1 Remove the top cover assembly (see “Removing the Top Cover Assembly” on page 141).
2 Place the top cover assembly top down on a cloth or other soft surface.
3 Disconnect the paper sense cable connector from the recorder adapter board.
4 Turn over the top cover and remove the screw holding the paper sensor to the RFI bracket.
Replacing the Paper Sensor Assembly
The procedure for replacing the paper sensor is a reversal of the removal procedure. Ensure that the
paper sensor cable is properly connected to the recorder adapter board.
Removing the Recorder Adapter Board
1 Remove the top cover assembly (see “Removing the Top Cover Assembly” on page 141).
2 Place the top cover assembly top down on a cloth or other soft surface.