12Specifications
Specifications for the accuracy of heart rate measurements are given in the Instructions for Use of the
fetal monitor.
Environmental Specifications
The monitor may not meet the given performance specifications if stored and used outside the
specified temperature and humidity ranges.
Monitor (M2702A/M2703A); Interface Cable for Avalon CTS (M2731-60001 and M2732-60001)
Temperature Range Operating Without battery option: 0°C—45°C (32°F—113°F)
With battery option/charging: 0°C—35°C (32°F—
95°F)
With battery option/fully charged: 0°C—40°C
(32°F—104°F)
Storage/Transportation -20°C—60°C (-4°F—140°F)
Humidity Range Operating <95% relative humidity @ 40°C (104°F)
Storage/Transportation <90% relative humidity @ 60°C (140°F)
Altitude Range Operating -500—3.000 m (-1.640—9.840 ft
Storage/Transportation -500—13.100 m (-1.640—43.000 ft)
Transducers (M2734A/M2734B/M2735A/M2736A/M2738A)
Temperature Range Operating 0°C—40°C (32°F—104°F)
Storage/Transportation -20°C—60°C (-4°F—140°F)
Humidity Range Operating <95% relative humidity @ 40°C (104°F)
Storage/Transportation <90% relative humidity @ 60°C (140°F)
Altitude Range Operating -500—3.000 m (-1.640—9.840 ft)
Storage/Transportation -500—13.100 m (-1.640—43.000 ft)
SpO
2
Sensors
Operating Temperature Range 0°C—37°C (32°F—98.6°F)