EasyManua.ls Logo

Philips Avalon FM20 - Status Log

Philips Avalon FM20
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Troubleshooting
123
Status Log
Many events that occur during start-up or regular monitoring are logged in the Status Log. The Status
Log can be cleared. Not all entries in the Status Log are errors. You can print the Status Log via the
Support Tool.
The
Status Log window shows logged events which caused a reboot of the monitor.
To enter the
Status Log window, select Main Setup, Revision. The following list opens up:
Status Log
Product
Appl SW
Config
Boot
Language
Settings
OB
FetRec
NBP (optional)
SpO(optional)
List of plugged parameters
Select
Status Log
The first column in the log identifies the event class ("C": caused a cold start, "H": caused a hot start,
"N": no restart, for information only). Column 3 and 4 identify the event source and event code.
Column 4 counts the number of occurrences of the event. The last column shows the time and date of
the last occurrence of the event.
The following pop-up keys overlay the SmartKeys:
Clear Stat Log
This key clears the currently displayed Status Log
M2703A
This key switches to the Monitor Revision Window
If an event occurs repeatedly, contact your Philips Service representative.

Table of Contents

Other manuals for Philips Avalon FM20

Related product manuals