10 Repair and Disassembly
156
12
Insert the new plastic hinge pins.
13 Do not apply excessive force when refitting the hinge pins. Turn the head of the hinge pin gently
with a small screwdriver. You will feel when the pin is seated correctly, and the head should be
flush with the surface.