12 Specifications
235
SpO
2
Performance Specifications
SpO
2
The specified accuracy is
the root-mean-square
(RMS) difference
between the measured
values and the reference
values
Range 0—100%
Accuracy Philips Reusable Sensors:
M1191A/B, M1191AL/BL, M1191ANL, M1192A, M1192AN = 2%
(70%—100%)
M1191T, M1192T, M1194A, M1194AN, M1196A, M1196T = 3%
(70%—100%)
Philips Disposable Sensors with M1943A(L):
M1131A, M1901B, M1903B, M1904B = 3% (70%—100%)
M1133A, M1134A = ±2% (70%—100%)
NellcorPB® Sensors with M1943A(L):
MAX-A, MAX-AL, MAX-P, MAX-N, D-25, D-20, N-25, OxiCliq A,
P, N = 3% (70%—100%)
Masimo Reusable Sensors® with LNOP MP12 or LNC MP10:
LNOP DC-I, LNOP DC-IP, LNOP YI, LNCS DC-I, LNCS DC-IP:
2% (70%—100%)
LNOP TC-I, LNCS TC-I: 3.5% (70%—100%)
Masimo Disposable Sensors® with LNOP MP12 or LNC
MP10:
LNOP Adt, LNOP Adtx, LNOP Pdt, LNOP Pdtx, LNCS Adtx,
LNCS Pdtx: 2% (70%—100%)
LNOP Neo-L, LNCS Neo-L: 3% (70%—100%)
Resolution 1%
Pulse Range 30—300 bpm
Accuracy ±2% or 1 bpm, whichever is greater
Resolution 1 bpm
Sensors Wavelength range 500—1000 nm.
Information about the wavelength range can be especially useful to
clinicians (for instance, when photodynamic therapy is performed).
Emitted Light
Energy
15mW
Pulse Oximeter Calibration Range 70%—100%
SpO
2
Alarm Specifications Range Adjustment Delay
SpO
2
50—100% 1% steps (0, 1, 2, 3,... 30) +
4 seconds
Desat 50—Low alarm limit 1% steps
Pulse 30—300 bpm 1 bpm steps (30—40 bpm)
5 bpm steps (40—300 bpm)
max. 14 seconds
Tachycardia Difference to high limit 0—50 bpm 5 bpm steps max. 14 seconds
Clamping at 150—300 bpm 5 bpm steps