12 Specifications
237
Physical Specifications
Rectal (ear + 1.16°C) 34.2—42.0 93.6—107.6
Caution: ASTM E1965-98 specifies 34.4°C—42.2°C (94°F—108°F)
Displayed Temperature Measurement Range
Mode Range °C Range °F
Ambient Temperature Range
Mode Range °C Range °F
Operating 10%—95% RH, non-
condensing
16.0—33.0 60.8—91.4
Storage up to 95% RH, non-
condensing
-25.0—55.0 -13.0—131.0
Caution: EN ISO 80601-2-56 specifies 16.0°C—35.0°C (60.8°F—95.0°F), 10%—95% RH, non-condensing
ASTM E1965-98 specifies 16.0°C—40.0°C (60.8°F—104.0°F), up to 95% RH, non-condensing
Storing the thermometer outside the specified temperature/humidity range might adversely affect measurement
accuracy. Check the calibration after storage in uncertain conditions.
Tympanic Temperature Alarm Specifications
Range 33.0°C—42.0°C (91.0°F—108.0°F)
Adjustment 0.5°C steps (33.0°C—35.0°C)
0.1°C steps (35.0°C—42.0°C)
1.0°F steps (91.0°F—95.0°F)
0.2°F steps (95.0°F—108.0°F)
Alarm delay System alarm delay (see “Fetal Monitors” on page 227).
Thermometer
Dimensions 190 mm x 43 mm x 55 mm (±3 mm)
Cable length 60 ±5 cm (spiral cable relaxed)
250 ±15 cm (spiral cable extended)
Weight (including cable) 180 ±10 g
Ingress protection classification IP 21
Base Station
Dimensions 205 mm x 65 mm x 75 mm (±3 mm)
Weight (excluding cable) 400 g ±10 g