7 Testing and Maintenance
57
Other Regular Tests and Tasks
Repairs
Upgrades
Tympanic Temperature performance testing Tympanic Temperature calibration
Service Event Test Block(s) Required - Complete these tests
Service Event Test Block(s) Required - Complete these tests
Visual Inspection Perform Visual Inspection
Transducer and Patient Module Testing See
Recorder Maintenance Regular cleaning and maintenance (see “Transducers and
Patient Modules; Functional Tests” on page 91)
Perform the recorder selftest (see “Fetal Recorder Selftest
Report” on page 98)
Service Event Test Block(s) Required - Complete these tests
Repairs when the monitor has been damaged by impact,
liquid ingression, fire, short circuit, or electrical surge
Perform Visual Inspection, Power On, all Safety tests, and
Full Performance Assurance tests
Repairs where the power supply, the mains socket or an
interface board is removed, or replaced, or the protective
earth ground connection is disrupted
Perform Visual Inspection, Power On, all Safety Tests,
and Basic Performance Assurance tests
Repairs where the NBP pump has been replaced Perform Visual Inspection, Power On Test, all Safety tests,
Basic Performance Assurance test, and NBP Performance
test and calibration
All repair events involving any of the following:
Noninvasive Blood Pressure connector, SpO
2
connector/
board or the red fetal sensor sockets
Perform Visual Inspection, Power On, Full Performance
tests, and all Safety tests
Service Event Test Block(s) Required - Complete these tests
Software upgrades Perform Visual Inspection, Power On test, and Basic
Performance Assurance test unless otherwise specified in
the Upgrade Installation Notes shipped with the upgrade
Hardware upgrades Perform Visual Inspection, Power On test, and Basic
Performance Assurance test unless otherwise specified in
the Upgrade Installation Notes shipped with the upgrade
Installation of interfaces or hardware where the power
supply, or parameter boards need to be removed
Perform Visual Inspection, Power On test, Basic
Performance tests, and all Safety tests
Noninvasive Blood Pressure hardware upgrade M2702AU
Option B71
Perform Visual Inspection, Power On, Noninvasive Blood
Pressure performance, and all Safety tests
SpO
2
hardware upgrade M2703AU Option B72 Perform Visual Inspection, Power On, SpO
2
performance, and all Safety tests
Noninvasive Blood Pressure/SpO
2
hardware upgrade
M2703AU Option B73
Perform Visual Inspection, Power On, Noninvasive Blood
Pressure/SpO
2
performance and all Safety tests