7 Testing and Maintenance
78
For any IT equipment (IEC 60950) operated in patient vicinity, ensure that the equipment leakage
current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure
compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is
required.