7 Testing and Maintenance
89
Noninvasive
Blood Pressure
Performance
Tests
Perform the
Accuracy Test -- see
“Accuracy Test” on
page 83.
X1 = value displayed by monitor
Difference <= 3 mmHg
PN:P/X1 or
PN:F/X1
Performance
Leakage Test -- see
“Leakage Test” on
page 84.
X2 = leakage test value
X2 < 6 mmHg
PN:P/X2 or
PN:F/X2
Performance
Linearity Test -- see
“Leakage Test” on
page 84.
X3 = value displayed by monitor
Difference <= 3 mmHg
PN:P/X3 or
PN:F/X3
Performance Valve
Test -- page
“Leakage Test” on
page 84.
X4 = value < 10 mmHg PN:P/X4 or
PN:F/X4
SpO
2
Performance
Test
Perform the SpO
2
Performance Test --
see “SpO2
Performance Test”
on page 85.
Value should be between 95%
and 100%
No reporting necessary
Safety (2) Perform Safety Test
(2): Sum of
Functional Earth
and Equipment
Leakage Current -
Normal Condition.
NC with mains cable:
Maximum leakage current (X1)
≤ 100 µA
S(1):P/X1 or
S(1):F/X1
Safety (3) Perform Safety Test
(3): Sum of
Functional Earth
and Equipment
Leakage Current -
Single Fault
Condition (Open
Earth).
SFC with mains cable:
Maximum leakage current (X2)
≤ 100 µA
S(2):P/X2 or
S(2):F/X2
Safety (4) Perform Safety Test
(4): Applied Part
Leakage Current -
Single Fault
Condition, mains on
applied part.
Maximum leakage current (X1)
≤ 50 µA
S(3):P/X1 or
S(3):F/X1
Test Block Test or Inspection
to be Performed
Expected Test Results Record the Results (mandatory for
Philips Personnel only)
What to Record Actual Results