Alarms, messages, and troubleshooting
Check Vent: Blower
Stalled
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(2)
Yes No No
Check Vent: Blower
Temperature High
Technical failure Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(33)
Yes No No
Check Vent: Cooling
Fan Speed Error
Overheating of ventila-
tor possible
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(36)
Yes No No
Check Vent: CPU
PCBA ADC Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(29)
Yes No No
Check Vent: Data Ac-
quisition PCBA ADC
Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(27)
Yes No No
Check Vent: Flash
File System Error
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(37)
Yes No No
Check Vent: Internal
Temperature High
CPU
Technical failure Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(30)
Yes No No
Check Vent: Internal
Temperature High
Daq
Technical failure Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(31)
Yes No No
Check Vent: Internal
Temperature High
Mtr
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(32)
Yes No No
Check Vent: Ma-
chine Pressure Sen-
sor Auto-Zero Failed
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(9)
Yes No No
Table 9-4: Check Vent alarm messages: summary and troubleshooting (continued)
Message Description Corrective Action