Alarms, messages, and troubleshooting
9-15
Check Vent: Ma-
chine Pressure Sen-
sor Calibration Data
Error
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(7)
Yes No No
Check Vent: Ma-
chine Pressure Sen-
sor Range Error
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(11)
Yes No No
Check Vent: Motor
Control PCBA ADC
Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(28)
Yes No No
Check Vent: O
2
Flow
Sensor Calibration
Data Error
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(14)
Yes No No
Check Vent: O
2
Pres-
sure Sensor Calibra-
tion Data Error
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(15)
Yes No No
Check Vent: O
2
Sup-
ply Pressure Sensor
Range Error
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(17)
Yes No No
Check Vent: OVP
Circuit Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(38)
Yes No No
Check Vent: Oxygen
Device Failed
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(16)
Yes No No
Check Vent: Primary
Alarm Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(4)
Yes No No
Check Vent: Pro-
gram CRC Test Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(1)
Yes No No
Check Vent: Proximal
Pressure Sensor
Auto-Zero Failed
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(10)
Yes No No
Check Vent: Proximal
Pressure Sensor Cali-
bration Data Error
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(8)
Yes No No
Table 9-4: Check Vent alarm messages: summary and troubleshooting (continued)
Message Description Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able