Chapter C - HELP & ERROR MESSAGES
Planmeca Proline XC X-ray unit C-3
ERROR MESSAGES
Technical Manual
Er 57
Exposure button activated when the unit is turned on.
Er 60
Generator processor power too low.
Er 61
Processor communication error.
Er 62
Limit sensor power missing.
Er 64
The radiation reaching the AEC sensor is very low or missing completely.
Er 65
The radiation reaching the AEC sensor is too high.
Er 70
Processor communication error.
Er 71
Generator processor program memory failure.
Er 72
Keyboard processor program memory failure.
Er 73
Keyboard processor program error.
Er 74
Keyboard processor operating improperly.
Er 75
Keyboard processor operating improperly.
Er 80
Keyboard processor EEPROM failure.
Er 81
Generator processor EEPROM failure.
Er 82
Keyboard processor configuration error.
Er 83
Generator processor configuration error.
Er 84
Tube power generator time out.
Er 85
PSU PCB internal voltage too low.
Er 86
Digital X-ray unit with two fixed sensor heads: Sensor head in wrong position.
Er 87
PSU PCB heat sink temperature too high.
Er 90
Keyboard processor EEPROM not activated.
Er 91
Keyboard processor stack overflow.
Er 99
Error code generation error.
ERROR CODE ERROR MESSAGE EXPLANATION