9 SYSTEM SETUP
User’s & installation manual Planmeca ProScanner 15
9 SYSTEM SETUP
9.1 Scanner placement
The scanner must be placed on a rigid and flat
desk or tabletop with at least 5 cm (2 inches)
free space on both sides, 10 cm (4 inches) on
back side and 15 cm (6 inches) on the front side
for imaging plate insertion.
Never place the scanner on the floor. Install
only in a location that is level and stable.
Installation in an unsuitable location can pro-
voke an accident, or deterioration in image
quality.
Sliding of the reader may result in internal
damage or misalignment of the optics. Exter-
nal vibration or shock during scanning may
affect image quality. The scanner must be
placed on a rigid, flat and reinforced desk or
tabletop.
Never install to locations with excessive
humidity or dust, subject to high tempera-
ture, subject to shaking or vibration,
exposed to considerable electrical or mag-
netic noise, or to any heath sources.
Do not place anything on top of the reader.
This equipment may be interfered with or
may interfere with electromagnetic or other
interferences. Assure a distance of minimum
1.0m between reader and neighbouring
equipment.