6 SYSTEM REQUIREMENTS
User’s & installation manual Planmeca ProScanner 9
6 SYSTEM REQUIREMENTS
The PC and other equipment connected to the
system must be:
• approved by local authorities: e.g. IEC-approved
(CE marked), UL / CSA approved
• located outside the patient area (more than 2m
(79 in.) from the X-ray unit)
• protectively earthed.
The connection of additional equipment to a
multiple portable socket-outlet must only be
possible by using a tool or be supplied via
separating transformer.
The multiple portable socket-outlets shall not be
placed on the floor.
Make sure that the system is protected with fire
wall and up-to-date anti-virus software. If
possible, isolate the system from office network.
Please note that these requirements are
recommendations. Not meeting them may lead
to degraded performance.