9 SYSTEM SETUP
User’s & installation manual Planmeca ProScanner 27
To set the IP-address manually, change the
CONNECTION parameter to MANUAL and fill in
the IP-address of the scanner to HOST
parameter.
//-----------------------------FireCR.ini-------------------
[SCANNER]
CONNECTION=MANUAL
[NETWORK]
HOST=192.168.1.134
//--------------------------------------------------------------
The configuration is not necessary if only one
scanner is used in the same network.
When configuring the IP-address manually, the
scanner DHCP mode must NOT be used.
N-to-N mode
When using N-to-N mode, it is not necessary to
configure the FireCR.ini file as the
CONNECTION parameter is by default set to
AUTO. The CONNECTION parameter cannot
be set to MANUAL in N-to-N mode.
However, make sure the Planmeca ProID
reader is attached to the PC before attempting
to connect from PC to the scanner in N-to-N
mode. If the reader is not connected, the
scanner will be connected in 1-to-1 ETH mode.
The device must be rebooted whenever the
mode is changed ie. if the scanner has been
used in 1-to-1 mode, a reboot is needed to
connect in N-to-N mode.