Loading...
Appendix A: Exposure value tables for Planmeca ProXA.2 Preprogrammed settings values
User’s & installation manual Planmeca ProScanner 53
Long cone 30 cm (12”)
INCISORS
PREMOLARS AND
CANINES
MOLARS
kV mA time kV mA time kV mA time
Adult Maxilla
60 8
0.16
63 8
0.2
63 8
0.25
Mandible 0.125 0.16 0.2
Child Maxilla
60 8
0.125
60 8
0.16
60 8
0.2
Mandible 0.1 0.125 0.16
OCCLUSAL ENDODONTIC BITE-WING
kV mA time kV mA time kV mA time
Adult
Maxilla
70 8 0,16 60 8 0.16 60 8 0.25
Mandible
Child
Maxilla
66
8
0,125 60 8 0.125 60 8 0.2
Mandible