E240 Overcurrent detected in collimator Y
motor which is connected to CAM
PCB J25 and controlled by MCM
module 5.
E241 Overcurrent detected in tube head
rotation motor which is connected to
CAM PCB J22 and controlled by
MCM module 6.
E242 Overcurrent detected in collimator
filter revolver motor which is
connected to CAM PCB J22 and
controlled by MCM module 6.
E243 Overcurrent detected in collimator
filter motor which is connected to
CAM PCB J20 and controlled by
MCM module 7.
E244 Overcurrent detected in collimator X2
left motor which is connected to CAM
PCB J20 and controlled by MCM
module 7.
Error
code
Explanation Comments
E245 Overcurrent Overcurrent detected in collimator Y2
lower motor which is connected to
CAM PCB J25 and controlled by
MCM module 7.
E246 Overcurrent detected in tube head
rotation motor which is connected to
CAM PCB J31 and controlled by
MCM module 8.
E251 Directional error Directional error at shoulder motor
which is connected to PSU PCB J7
and controlled by MCM module 1.
The motor drives the
mechanism into the wrong
position. Check the sensors
and the cabling.
E252 Directional error at elbow motor which
is connected to PSU PCB J7 and
controlled by MCM module 2.
The motor drives the
mechanism into the wrong
position. Check the sensors
and the cabling.
E256 Directional error at c-arm motor which
is connected to CAM PCB J21 and
controlled by MCM module 4.
The motor drives the
mechanism into the wrong
direction. Check the
sensors and the cabling.
E267 Patient support error Patient support moved during the
patient support locking mechanism.
3 Error messages