E291 Collimator Collimator right X self test failed.
E292 Collimator left X self test failed.
E293 Collimator upper Y self test failed.
E294 Collimator lower Y self test failed.
E295 Collimator filter revolver self test
failed.
E296 Collimator X collision detected.
E297 Collimator Y collision detected.
E298 Tube head rotation Tube head rotation self test failed
E299 Z lock Z lock test failed
X-ray generation related errors (E3xx)
The following table lists the X-ray generation related errors.
Error
code
Explanation Comments
E301 Filament voltage Filament voltage missing
completely.
Exposure not possible. Check
the cabling, interfaces and
PSU- PCB and tube head.
E302 Filament voltage too low during
preheat.
Exposure not possible. Check
the cabling, interfaces and
PSU- PCB and tube head.
E303 Filament voltage too high during
preheat.
Exposure not possible. Check
the tube head and the PSU-
PCB.
Error
code
Explanation Comments
E311 Tube voltage Tube voltage missing completely. Exposure interrupted.
E312 Tube voltage too low. Exposure not possible. Check
the tube head and the PSU
PCB.
E313 Tube voltage too high. Exposure not possible. Check
the prefilament values are
correct or perform the
prefilament calibration again.
E317 Tube voltage overshot suddenly
without arching.
Exposure interrupted. Check
the interface, cabling and the
PSU PCB.
E318 Tube voltage Tube mA overshot suddenly
without arching.
Exposure interrupted. Check
the functionality of the tube
head and the PSU PCB.
E319 Tube voltage Tube pulse length exceeded the
requested value.
3 Error messages
Technical manual Planmeca Viso 41