PDR-7000 Operation Manual
To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
Only operate the instrument with the power supply indicated on the
rating plate. Otherwise, it may result in fire or electric shock.
Be sure to turn OFF the power switch before connecting or
disconnecting the cables. Also, do not handle them with wet hands.
Otherwise, you may get an electric shock that may result in death
or serious injury.
Never disassemble or modify this instrument because it may result
in fire or electric shock. Also, since this instrument incorporates
high-voltage parts and other hazardous parts, touching them may
cause death or serious injury.
Should any of the following occur, immediately turn OFF the power
switch, unplug the power cable from the AC outlet, and contact the
dealer or the agent who/where you purchase this instrument.
When there is smoke, strange odor or abnormal sound.
When liquid has been spilled into the instrument or a metal
object has entered through an opening.
When the product has been dropped or its housing damaged.
This instrument is shipped with a grounding type power cable. To
reduce the risk of electric shock, always plug the cable into a
grounded power outlet.
Wipe the forehead rest with ethanol or glutaraldehyde solution to
disinfect it each time a different examinee uses it, in order to
prevent infection.
Do not use the device near, on, or under other electronic equipment
to avoid electromagnetic interference with the operation of the
device.
Do not use the device in the same room with other equipment such
as life-support equipment, other equipment that has major affects
on the life of patient an results of treatment , or other measurement
or treatment equipment that involve small electric current.
Do not use the device simultaneously with portable and mobile
radio frequency communication systems because it may have an
adverse effect on operation of the device.
Do not use cables and accessories that are not specified for the
device because that may increase the emission of electromagnetic
waves from the device or the system and decrease the immunity of
the device to electromagnetic disturbance.
Do not place the multiple socket-outlet for PDR-7000 system on the
floor in order to prevent liquid penetration and damage to the
product.
PDR-7000 system shall not be connected with additional multiple
socket-outlets or extension cords in addition to a designated single
multiple socket-outlet.
Maximum permissible load of each socket-outlet used for the PDR-
7000 system, shall not be less than 100VA.
- v -