To pin VDE cable Contact Relay
1 yellow NO (normally open) Other relay
4 green C (common) Other relay
3 white NO (normally open) Pacing relay
2 brown C (common) Pacing relay
9.3.3 Output "Current output plus relay" (identity code C + D + E)
A relay combined with a current output can be optionally ordered. The
relay switches either as a fault indicating relay when there is a fault with
the pump and with "Liquid level low, 1st stage" warning alert and "Liquid
level low 2nd stage" fault alert or is used as a pacing relay.
The behaviour is programmed ex factory. If another switching function is
required, the pump can be reprogrammed in the
‘Relay’
menu.
In the
‘ANALOG OUTPUT’
menu, the variable to be signalled by the cur‐
rent output can be selected.
The current output plus relay can be retrofitted and operates once it is
plugged into the board.
Electrical interface
for current output
Data Value Unit
Open circuit voltage: 8 V
Current range: 4 ... 20 mA
Ripple, max.: 80 μA ss
Ripple, max.: 250 Ω
for semiconductor switch ("relay"):
Data Value Unit
Residual voltage max. at I
off max
= 1 µA 0.4 V
Maximum current 100 mA
Maximum voltage 24 VDC
Pacing pulse duration, approx. 100 ms
To pin VDE cable Contact Relay
1 yellow "+" Current
output
4 green "-" Current
output
3 white NC (normally closed) or
NO (normally open)
Relay
2 brown C (common) Relay
Identity code 4 + 5, 8 + 9, A + B
Fig. 21: Cable conductor assignments
Fig. 22: Pump pin assignments
Identity code C + D + E
Fig. 23: Cable conductor assignments
Electrical installation
37