Sensor Connector Cell con‐
stant
ZK (1/cm)
T-correc‐
tion ele‐
ment
Max.
temp.
(°C)
Measuring
range κ
min
(Unit)
Measuring
range κ
max
(Unit)
LFTK1-DE DIN 4-pin 1.00 Pt1000 80 0.01 mS/
cm
20 mS/cm
LFTK1-1/2 DIN 4-pin 1.00 Pt1000 80 0.01 mS/
cm
20 mS/cm
LF1-DE DIN 4-pin 1.00 - 80 0.01 mS/
cm
20 mS/cm
LFT1-DE DIN 4-pin 1.00 Pt100 80 0.01 mS/
cm
20 mS/cm
LFT1-1/2 DIN 4-pin 1.00 Pt100 80 0.01 mS/
cm
20 mS/cm
LMP01 DIN 4-pin 0.10 Pt100 70 0.1 uS/cm 500 uS/cm
LMP01-
HT
DIN 4-pin 0.10 Pt100 120 0.1 uS/cm 500 uS/cm
LMP01-TA Fixed
cable
0.34mm
2
,
5 m,
shielded
0.10 Pt100 70 0.1 uS/cm 500 uS/cm
LMP001 DIN 4-pin 0.01 Pt100 70 0.01 uS/
cm
50 uS/cm
LMP001-
HT
DIN 4-pin 0.01 Pt100 120 0.01 uS/
cm
50 uS/cm
LM1 DIN 4-pin 1.00 - 70 0.1 mS/cm 20 mS/cm
LM1-TA Fixed
cable 0.34
mm
2
, 5 m,
shielded
1.00 - 70 0.1 mS/cm 20 mS/cm
LMP1 DIN 4-pin 1.00 Pt100 70 0.1 mS/cm 20 mS/cm
LMP1-HT DIN 4-pin 1.00 Pt100 120 0.1 mS/cm 20 mS/cm
Sensor connection
41