Sensor status
Display Meaning Status
[OK]
In order Cell constant = 0.005 / 15.0
[WRN]
Warning none
[ERR]
Error Cell constant < 0.005 or cell constant > 15
9.1.2 Calibration of the temperature coefficient
Conductivity sensors with temperature element
You can only calibrate the temperature coefficient with conductivity sensors with a
temperature elements, because it is impossible to calculate the temperature coeffi‐
cient without measuring the temperature.
Temperature change
It is recommended that the temperature is changed by no more than 0.5 °C per
minute, or with a temperature change of e.g.10°C you will need to wait for a min‐
imum of 20 minutes before calibration.
Operating menus
60