Response of the device due to the error messages
Error message Control
mode
mA measure‐
ment output
mA correc‐
tion output
Limit
value
relay
Suppressed
during user
calibration
[RANGE↧]
Basic load Error current - - yes
[RANGE↥]
Basic load Error current - - yes
[T RANGE↧]
Basic load Error current Error cur‐
rent
- yes
[T RANGEt]
Basic load Error current Error cur‐
rent
- yes
[CALERROR]
- - - - yes
[LOW ZERO]
- - - - yes
[CHECKTIME]
Basic load Error current - - no
[mA RANGE↥]
- - - - no
[mA RANGE↧]
- - - - no
[LIMIT↥]
- - -
activation
1
yes
[LIMIT↧]
- - -
activation
1
yes
[T LIMIT↥]
- - - - no
[T LIMIT↧]
- - - - no
[LIMIT ERR]
Stop Error current - - yes
[TLIMITERR]
Stop Error current Error cur‐
rent
- no
[NOCAL]
- - - - yes
[CONTACTIN]
- - - - no
[TDS↥]
- - - - no
[SAL↥]
- - - - no
[INPUT↥]
Basic load Error current - - no
1)
If the functionality of the limit value relay is activated and the switch-on delay is over‐
come.
Maintenance
92