Battery level drops too quickly
The drop-in battery level is not necessarily linear across time. The battery
level may also fluctuate when a sensor is connected or disconnected.
However, the operating duration should still be more than 4 hours on a
single full charge under normal use (i.e. with screen dimming and no USB
connection).
POSSIBLE CAUSES
ACTION
Low ambient temperature
If the device is left or operated for prolonged
period under an ambient temperature of below 0
degree Celsius, the total operating duration may
be reduced.
Damaged battery
If the battery is less than 6 months old and the
normal operating duration is significantly
shorter than 4 hours at room temperature, then
pre-mature battery damage is indicated.
Abnormal power
consumption
The sensor draws majority of its power during
operation. Verify that there is no abnormal
power consumption caused by a faulty sensor.