EasyManuals Logo

QuadTech 1730T User Manual

Default Icon
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
Page 72 of 110 150821/A4 Interface
Table 3-4 lists the IEEE-488 interface commands the 1730T instrument accepts to set or query a
parameter value. Paragraphs 3.1.3 through 3.1.5 detail the command function, format, return
value and description.
Table 3-4: IEEE-488 Commands
Command Name Function Output Format
*CLS Clear Status Clear standard event status
register. Clear status bit group
register except for bit 4 (MAV)
*ESE Event Status Enable Enable standard event status
register value.
0 – 255
*ESE? Event Status Enable Query standard event status of
device enable register
0 – 255
*ESR? Event Status Register Query standard event register
value of device. After this
command, the standard register is
cleared to 0.
0 – 255
*IDN? Identification Query/Read basic device data.
(A comma separates the
identification fields.)
4 ID:
Manufacturer,
Device Model,
Serial Number,
Firmware Version
*OPC Operation Complete Operation is complete. 0
*OPC? Operation Complete Query operation complete. 1
*RST Reset Reset Device.
*SRE Service Request Enable Enable service request register
value.
0 – 255
*SRE? Service Request Enable Query/Read service request
register value.
0 – 255
*TRG Trigger Bus Trigger the 1730T instrument
*TST? Self Test Perform self test & report error 0 = no error
1 = RAM1
2 = RAM2
4 = EEPROM
8 = FPGA
16 = Cal Data
32 = Main Board
64 = Key Control
*SAV Save Save current status to memory. 1 – 50
*RCL Recall Recall saved status from memory. 1 – 50

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the QuadTech 1730T and is the answer not in the manual?

QuadTech 1730T Specifications

General IconGeneral
BrandQuadTech
Model1730T
CategoryMeasuring Instruments
LanguageEnglish