Operation 150821/A4 Page 35 of 110
2.3.16 LINK 1320
It is possible to link the 1730T instrument to the QuadTech 1320 Bias Current Source. The 1320
instrument is capable of producing 20A DC bias current. The 1730T instrument default setting
is OFF. To change the LINK 1320 press [SYSTEM SETUP], [SYSTEM CONFIG] and the
down arrow [⇓] until the box next to LINK 1320 is highlighted, then press [F1] = ON or [F2] =
OFF.
<SYSTEM CONFIG>
ON
OFF
0AhGPIB EOS CODE:
LINK 1320 :
TRIGGER EDGE:
TRIGGER DELAY:
GPIB ADDRESS:
OFF
FALLING
0000mS
17
F2
Select link to 1320 OFF
Default: OFF
Select link to 1320 ON
F1
Go to VM / IM Display
Go to TRIGGER Edge
2.3.17 HARM CHECK
To have the 1730T run an instrument status check when the instrument is powered ON, select
HARM CHECK = ON. HARM CHECK can be set to ON or OFF and the instrument default
value is ON. Board status is checked for the MAIN board, RAM1, Key Control, GPIB &
Handler, RAM TRANSIT, EEPROM, FPGA Download and Calibration Data.
F2
Select Harm Check OFF
Default: ON
Select Harm Check ON
F1
Go to VM / IM Display
Go to LINK 1320
<SYSTEM CONFIG>
ON
OFF
0AhGPIB EOS CODE:
LINK 1320 :
TRIGGER EDGE:
TRIGGER DELAY:
GPIB ADDRESS:
OFF
FALLING
0000mS
17
HARM CHECK :
ON
QuadTech, INC
http://www.quadtech.com
COPYRIGHT (c) JAN 2006, VER 1.20
1730T LCR Meter
PASSEDMAIN BOARD
RAM TRANSIT
GPIB 17 & HANDLER
KEY CONTROL
RAM1
PASSED
PASSED
PASSED
PASSED
EEPROM CHECK PASSED
FPGA DOWNLOAD PASSED
CALIBRATION DATA PASSED
About
Flash Screen
Harm Check
Flash Screen
l
Press Power ON