EM510 SENSe Subsystem :PSCan
4065.7763.32-01.00 5.90
. :PSCan
The PSCan system checks the frequency function of the unit in case a panorama scan was initiated by
SENSe:FREQuency:MODE. Each scan is started only by INITiate[:IMMediate].
. . :COUNt <numeric_value>|MINimum|MAXimum|INFinity
Output of a number of PSCan cycles.
Parameters:
<numeric_value> number of cycles
MINimum|MAXimum minimal/maximal number
INFinity infinite number
*RST state:
INFinity
Example:
PSCan:COUNt 100
. . :STEP<numeric_value>|MINimum|MAXimum| UP|DOWN
Setting of the channel raster and the resolution bandwidth correspondingly.
Parameters:
<numeric_value> the following discrete values can be set: 125 Hz, 250 Hz,
500 Hz, 625 Hz, 1.25 kHz, 2.5 kHz, 3.125 kHz,
6.25 kHz, 12.5 kHz, 25kHz, 50 kHz, 100kHz
MINimum|MAXimum setting of the smallest/widest resolution bandwidth
UP | DOWN setting of the next smallest/widest resolution bandwidth
*RST state:
100 kHz
Example:
PSCan:STEP 25 kHz
. . . ? [MINimum|MAXimum]
Query about the resolution bandwidth at the panorama scan.
Parameters:
none query about the current resolution bandwidth
MINimum|MAXimum query about the smallest/widest resolution bandwidth
Result:
Resolution bandwidth in Hz
Example:
PSCan:STEP? 25000