R&S EM510 - Firmware Update Messages and Indication

R&S EM510
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EM510 Preparation for Use
4065.7763.32-01.00 4.6
4.4.4 Messages and Indication
4.4.4.1 Update Messages
UPD32 displays the following messages during the update procedure:
STARTING COMMUNICATION FOR TARGET (ADR:00 90 B8 10 01 14)
Load boot code....
Erasing...
Loading program code...
Calculating checksum...
UPDATE COMPLETE
4.4.4.2 Indication on the EM510 front panel
The lighted ACCE LED on the front panel of the EM510 shows that the update is in progress.
After the update the ACCE LED is switched off and the FAIL LED is switched on.
If no problem is detected, the EM510 application starts automatically and switches off the FAIL LED.
If a problem appears, the FAIL LED toggles the error codes.
Error Code Definition Note
0 ERROR_NO_ERROR No error
1 ERROR_CPLD_NOT_PROGRAMMED Internal
2 ERROR_EEPROM_NO_IDENT_BLOCK Internal
3 ERROR_EEPROM_NO_SERIAL_NUMBER Internal
4 ERROR_EEPROM_NO_ETH_ADDR Internal
5 ERROR_WRONG_LOCATED Internal
6 ERROR_NO_APPLICATION There is no application in the Flash
EEprom