5.5.5 Resetting Values of the Status Reporting System
Table 5-9 comprises the different commands and events causing the status reporting system to be
reset. None of the commands, except for *RST, influences the functional device settings. In particular,
DCL does not change the device settings.
Table 5-9: Resetting device functions
Event
Switching on supply
voltage DCL, SDC
(Device Clear,
Selected Device
Clear)
*RST STATus:PRESet *CLS
Effect
Clear STB, ESR yes yes
Clear SRE, ESE yes
Clear PPE yes
Clear EVENTt sections of
the registers
yes
yes
Clear ENABle section of all
OPERation and
QUEStionable registers,
Fill ENABle sections of all
other registers with "1".
yes
yes
Fill PTRansition sections
with "1" ,
Clear NTRansition sections
yes
yes
Clear error queue yes
yes
Clear output buffer yes yes 1) 1) 1)
Clear command processing
and input buffer
yes yes
yes
1) The first command in a line, i.e. immediately following a <PROGRAM MESSAGE TERMINATOR>,
clears the output buffer.