EM510 MEASure Subsystem
4065.7763.32-01.00 5.40
. :MODE?
Query about the set measuring mode.
Parameters:
none
Result:
CONT continuous measurement
PER periodic measurement
Example:
MEASure:MODE? -> PER
. :TIME <numeric_value>|MINimum|MAXimum|DEFault
Setting the duration for all measurement functions.
Note:
The user is responsible for setting a useful measuring time. Time spans which are too short lead to
faulty measurement results.
Parameters:
<numeric_value> time span in seconds
MINimum|MAXimum min/max time span
DEFault default time span
*RST state:
DEFault
Example:
MEASure:TIME 200 ms
. :TIME? [MINimum|MAXimum]
Query about the set measuring time.
Parameters:
none queries the current time span
MINimum|MAXimum queries the min/max time span
Result:
Time span in seconds; the default time span is indicated by DEF
Example:
MEASure:TIME? -> 0.2