EasyManua.ls Logo

R&S EM510 - CALCulate Subsystem: IF-Panorama Averaging

R&S EM510
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CALCulate Subsystem EM510
5.19 4065.7763.32-01.00
5.3.4 CALCulate Subsystem
CALCulate
. :IFPan
. . :AVERage
. . . :TYPE MINimum|MAXimum|SCALar|OFF
Setting of the averaging procedure for the IF-panorama data
Parameters:
MINimum MIN hold function is on
MAXimum MAX hold function is on
SCALar AVG averaging function is on
OFF Switching on the Clear Write function
Note:
For the averaging procedure the averaging time and measuring time can be set by commands
MEASure:TIME. The IF-panorama is also used for bandwidth measurement. With measure mode
periodic the detectors are discharged in the cycle of the measure time. Therefore the display of the IF
panorama “pulses” in the cycle of the measure time.
*RST state:
OFF
Example:
CALCulate:IFPan:AVERage:TYPE MAXimum
. . . :TYPE?
Query about the averaging process of the IF-panorama data
Result:
MIN MIN hold process is running
MAX MAX hold process is running
SCAL AVG averaging process is running
OFF Switching on the Clear Write function
Example:
CALCulate:IFPan:AVERage:TYPE? -> MAX
. . :CLEar
Restart of the MIN or MAX hold function for the IF-panorama data
Parameters:
none
Example:
CALCulate:IFPan:CLEar

Table of Contents