EM510 Installation and Cabling
4065.7763.32-01.00 3.2
3.3 Bench Operation
Attention: Do not expose the EM510 to humidity.
Leave at least 50 mm of empty space along both side panels with ventilation slots in order
to ensure proper cooling function.
There are no special requirements for bench operation.
To facilitate access to the front panel elements, we suggest that you raise the front of the EM510 by
deploying the unit's feet.
3.4 Rack Mounting
Attention: The EM510 should be used in an area where the ambient temperature does not
exceed –10 to +55 °C.
The EM510 is fan cooled and must be installed with sufficient space on the sides to permit
a free flow of air. Make sure that hot air can escape freely. To ensure sufficient cooling, do
not attach telescopic rails to the sides of the unit.
If the rack is exposed to high ambient temperatures, sufficient ventilation must be ensured for the rack.
Note: Do not remove the enclosure when rack mounting the EM510.
Place the EM510 onto the guide rails of the rack and insert it into the rack. Use the four screws of
the front mounting bracket to fasten the unit to the rack.
3.5 Cabling
Cabling always depends on the particular application involved.
Connect the EM510 observing the following sections and instructions for use.
For further connections not mentioned in the following sections refer to the Interfaces Description.