DIAGnostic Subsystem EM510
5.27 4065.7763.32-01.00
. :TEMPerature[<numeric_suffix>]?
Output of the temperature at various measurement points of the device.
Parameters:
The measurement point (1 to 5) is selected through the <numeric suffix>.
1 or none <numeric suffix> TEMP_LOCAL temperature on the mainboard
2 TEMP_FPGA temperature of the FPGA
3 TEMP_PPC temperature of the Power PC
4 TEMP_ADSECT temperature of AD converter section
5 TEMP_PRESEL temperature in the preselection
Result:
Current temperature in °C at the selected test point.
Example:
DIAGnostic:TEMPerature3? -> 37
. :TPOint[<numeric_suffix>]? <module>
Output of test point voltage of a module. The test point (1 to n
max
) on the particular module is selected
via <numeric suffix>:
Parameters:
<module> information relating to module to be queried:
V1 preselection n
max
= 3
MB mainboard n
max
= 16
Result:
Current test point voltage in mV
Example:
DIAGnostic:TPOint1? V1 -> 1448