SENSe Subsystem EM510
5.59 4065.7763.32-01.00
. : CORRection:ATTenuation?
Read out of the current total attenuation in the signal path from antenna input to IQ output. The
attenuation depends on the current receiver frequency and the current temperature.
The absolute signal power can be calculated by adding this value to the IQ data level dB full scale. The
automatic attenuation selection and the automatic gain control has to be switched off in this case.
Result:
Attenuation in dB
Example:
CORRection:ATTenuation? -> 105