EasyManua.ls Logo

Rigaku MFM65 - Lock-Out; Tag-Out

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
ME12058B
Maintenance Section
2.4 Lock-out / Tag-out
WARNING
When electrical power off and isolating the cooling water of facility, strictly follow
lock-out/tag-out of the instrument as follows.
In the daily operation, the lock-out/tag-out of the cooling water is normally
unnecessary.
(1) The main breaker lock-out handle
1. Pull down the lock-out handle.
2. Set the tag to the lock-out handle.
3. Lock the tag.
(2) The cooling water lock-out
1. Open the lower panel on the left side.
2. Remove the internal cover.
.
The
p
ower box on the rear side
Fig. 2-4-1 Lock-out Handle
Fig. 2-4-2 Water Lock-out
4. Lock-out the bulbs.
3. Close the bulbs

Table of Contents