80
ME12058B
Maintenance Section
15.3 Beam Diagnosis
Troubles Causes and remedies
The direct beam intensity and
position are not as specified.
If the intensity is low, the X-ray optics may be out of
adjustment.
If the direct beam position is different, the X-ray optics may
be out of adjustment.
NOTE:
Contact our service personnel in charge for
readjustment of the X-ray optics.
If the intensity does not improve after the
readjustment, the X-ray tube may have
reached the end of service life. Check to see
if the X-ray tube needs to be replaced.
NOTE:
Contact our service personnel in charge for
readjustment of the X-ray optics.