Maintenance Section
Troubles Causes and remedies
Check to make sure that the software screen does not show an
FCL alarm.
Shut down the PC, turn off the main breaker lockout handle
of the equipment, and restart the equipment.
(Refer to “2.4 Lock-out/Tag-out”.)
If this alarm is generated again, the filament in the X-ray tube
may be disconnected.
If the software screen shows another alarm or if the cause of
the problem is unknown, shut down the PC, turn off the main
breaker lockout handle of the equipment, then restart the
equipment. (Refer to “2.4 Lock-out/Tag-out”.)
CAUTION
When the X-ray tube is replaced, contact
our service personnel in charge.