− 88 −
ME12058B
Maintenance Section
(2) List of maintenance parts
Items Models Suppliers Remarks
Multi-layers mirror for Cu C086-13-1
Rigaku
8
Multi-layers mirror for Mo C086-14-1
Rigaku
8
UPS Battery BS10A-H24/2.0L
Nipron
8
Robot Hand C093-25-1
Rigaku
8
X-ray detector (APD) A989-001-3-4
Rigaku
8
X-ray detector (SDD) C093-24-1
Radiant
8
X-ray detector (SC70) Cat.No.5738E401
Radiant
8
Load port TAS300
TDK
8
Keyboard PD-KB200B/U
Rigaku
Trackball H56-5600 White
Rigaku
Joystick 30JEK-ZU-06R3
Rigaku
LCD LCD-AD174CB
Rigaku
DC power supply AC4-000HH-00
Rigaku
8
FAN (bottom) T-MDS1451-24L-G
Rigaku
8
FAN (driver) T-MD825B-24L
Rigaku
8
FAN (rear) T-MD825-24L
Rigaku
8
Surge arrester RAV-781 BXZ-2A
Rigaku
8
Fuse (RCD/XGC) NF60-NR5A
Rigaku
8
Fuse (AFC) T1L250V
Rigaku
8
Fuse (Power Box) FNQ-R-15
Rigaku
8
NOTE:
Items of 8 mark should be exchanged by Rigaku service personnel.
Contact to Rigaku service center for exchange.
WARNING
Metal Beryllium foil is used for the windows of X-ray detector APD and SDD.
Solid Beryllium usually has no influence on human health, but evidence exists
that the personnel who inhales the beryllium powder or vapor risks a possibility
of serious injury or death. Do not scratch, polish, and sweep with neither
chemical nor bake up to temperature so as to generate vapor or powder. In case
that beryllium window is ruptured, Personal Protective Equipment must be
used and collect as much chips as possible. Then dispose them according to
local regulations. If difficulty in local disposal exists, please gather the pieces
of fractured Beryllium window up carefully and send back them together with
X-ray detector back to the manufacturer.
(Refer to the attached MSDS.)