Remote Commands for the I/Q Analyzer
R&S
®
FSWP
115User Manual 1177.5856.02 ─ 07
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 15
Switches on 15 dB preamplification.
Manual operation: See " Preamplifier " on page 50
INPut<ip>:GAIN[:VALue]
<Gain>
This command selects the gain if the preamplifier is activated (INP:GAIN:STAT ON,
see INPut<ip>:GAIN:STATe on page 114).
The command requires the additional preamplifier hardware option.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Gain> 15 dB | 30 dB
The availability of gain levels depends on the model of the
R&S FSWP.
R&S FSWP8/26: 15dB and 30 dB
R&S FSWP50: 30 dB
Default unit: DB
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation: See " Preamplifier " on page 50
8.6.6 Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]....................................................................116
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE............................................... 116
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE..............................116
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition....................................................117
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing........................................117
Configuring I/Q Analyzer Measurements