Remote Commands for the I/Q Analyzer
R&S
®
FSWP
92User Manual 1177.5856.02 ─ 07
INITiate:SEQuencer:REFResh[:ALL]
This function is only available if the Sequencer is deactivated (SYSTem:SEQuencer
SYST:SEQ:OFF) and only in MSRA mode.
The data in the capture buffer is re-evaluated by all active MSRA slave applications.
Example:
SYST:SEQ:OFF
Deactivates the scheduler
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a new data measurement and waits for the end of the
sweep.
INIT:SEQ:REFR
Refreshes the display for all channels.
[SENSe:]AVERage:COUNt <AverageCount>
TRACe:IQ:AVERage:COUNt <NumberSets>
This command defines the number of I/Q data sets that the averaging is based on.
Parameters:
<NumberSets> Range: 0 to 32767
*RST: 0
Example:
TRAC:IQ ON
Switches on acquisition of I/Q data.
TRAC:IQ:AVER ON
Enables averaging of the I/Q measurement data
TRAC:IQ:AVER:COUN 10
Selects averaging over 10 data sets
TRAC:IQ:DATA?
Starts the measurement and reads out the averaged data.
Manual operation: See " Sweep/Average Count " on page 64
[SENSe:]AVERage<n>[:STATe<t>] <State>
TRACe:IQ:AVERage[:STATe] <State>
This command turns averaging of the I/Q data on and off.
Before you can use the command you have to turn the I/Q data acquisition on with
TRACe:IQ[:STATe].
If averaging is on, the maximum amount of I/Q data that can be recorded is 512kS
(524288 samples).
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Performing Measurements